HILIC-MS impurity profiling of therapeutic PS- oligonucleotides

CAST scientist Luca Tutiš together with Govert Somsen and Andrea Gargano as well as AstraZeneca, Novartis and GSK have just published a manuscript describing a novel LC-MS method for impurity profiling of therapeutic phosphorothioated oligonucleotides.
Facebook
Twitter
Pinterest
LinkedIn

Ion-Pairing Hydrophilic Interaction Chromatography for Impurity Profiling of Therapeutic Phosphorothioated Oligonucleotides

Oligonucleotides are short strands of synthetic DNA or RNA that are synthesized via a solid-phase synthesis, in which numerous of closely-related impurities are generated. Ion-pairing reversed-phase liquid chromatography (IP-RPLC), anion exchange chromatography (AEX), and hydrophilic interaction chromatography (HILIC) are often used to profile these impurities, which allows for good separation of impurities comprising different number of nucleotides as the full-length product (FLP). However, impurities comprising the same number of nucleotides as the FLP are often not separated. Therefore, ion-paring HILIC (IP-HILIC) was explored as an alternative separation mode to overcome these challenges.

Key points:

  • Changed selectivity: by adding ion-pairing reagents (IPRs) to the HILIC eluent, the relative contribution of the highly polar phosphate moieties on HILIC retention is reduced and, thereby, increasing the relative contribution of the nucleobase composition and conjugated groups.
  • Suppressed diastereomer separation: Phosphorothioation of the phosphate groups results in the formation of diastereomers, with 2n possible diastereomers (n = phosphorothioate groups). IPRs in the HILIC eluent reduced diastereomer separation, leading to sharper peaks.
  • Separation of same-length impurities: IP-HILIC shows increased separation of impurities comprising of the same number of nucleotides as the FLP, such as deaminated products that differ less than 1 Da from the FLP. This is noteworthy as no other MS-compatible, one-dimensional LC separation can achieve this.

Figure 1: IP-HILIC-MS total and extracted ion chromatograms of GalNAc-conjugated oligonucleotides (top) and non-conjugated oligonucleotides (bottom) and the mass spectra of peaks A-D (A & C: FLPs, B & D: deaminated products)

The developed IP-HILIC method shows great potential as a screening method for quality control. The work is published in Analytical Chemistry and can be found with the following link: https://pubs.acs.org/doi/full/10.1021/acs.analchem.5c01407

 

Related Posts

Novel sequencing approach enables detailed characterization of PLGA co-polymers

CAST Scientst Masashi Serizawas was published in Macromolecules. Masashi presents a ‘polymer sequencing’ approach for analyzing the composition of PLGA copolymers in great detail. The method consists of controlled chemical degradation followed by reversed-phase liquid chromatography and high-resolution mass spectrometry (RPLC-HRMS) to arrive at an accurate characterisation of block-length distribution. The work provides a robust framework for the rational design of sequence-controlled PLGA for drug delivery and other advanced applications.

Microscale native SEC-HRMS to characterize immunoconjugates

In this study, CAST scientists Annika van der Zon et al., in collaboration with Bram Weijers and Danielle Vlugts (Amsterdam UMC), introduce a rapid, microscale aqueous size-exclusion chromatography–mass spectrometry (SEC-MS) method to measure chelator to antibody ratio in immunoconjugates.

Size-Resolved Surface Charge Analysis of Polymer Nanoparticles: From Fundamental Measurement to Collaborative Innovation

The study addresses size-resolved surface charge density (SCD) analysis of polymer nanoparticles (PNPs), revealing the significant variability of SCD that traditional methods overlook. By employing advanced capillary zone electrophoresis techniques, researchers can gain detailed insights into nanoparticle performance, benefiting applications in coatings, biomedical, and industrial contexts through collaborative innovations.